Detectable Species for Quantification
- Free radicals
- Paramagnetic transition metal ions
- Metals
- Crystal lattice (defects)
Samples
- Aggregation from liquid to solid,
e. g. liquids, solutions, suspensions, blood, cells, paste, vessels,
tissue, powder, solid compound
Temperature of the Sample
- Resonator temperature (32 -
34°C)
- - 196 °C
(Accessories: Dewar, liquid nitrogen)
- - 170 to
+ 200 °C for capillaries and sample tubes of
outer diameter 3 mm (Accessory: Temperature Controller TC H02)
Detection Limit
- 25 nM TEMPOL
in 50 mM phosphate buffer, pH 7,4 *
- 2 x 10
Spins/0,1 mT
* Sample vessel
capillary, microwave power 20 mW, modulation amplitude
0,2 mT, scan time 4 times 60 s, time constant
100 ms
Packages for Special Applications
- Alanine dosimetry
- NO-Fe(II)DETC
- Reactive oxygen species (ROS)
Packages for Automation
- Flow through system, Multi-Channel
- Flow through system, One-Channel
|
 |

Sample Vessels
- Capillaries, sample tubes (outer
diameter 3 - 6 mm)
- Flat cells (Width of the flat
section 5 or 7 mm)
- Tissue cell
- Dewar
- Special cuvettes
1 µM TEMPOL, 50 mM
phosphate buffer pH 7.4*

Accessories
- Temperature Controller TC-H03
- Auto sampler
- General accessories
- Glassware
- Software
|
- Spectrometer
- Guidance Tube F 120/1.5 for
capillaries
- 250 pieces 50 µl
capillaries
- Sealing kit for capillaries
- Standard-Software
- MiniScope
Control
(control software):
- Adjust of the measurement
parameters
- Automatic or manual tuning
- Recording of Spectra
- Saving of data
- 4096 data points per spectrum,
etc.
|
|
- Analysis
(calculation software):
- Amplitude calculation
- Integration
- Baseline correction
- Filtering
- g-value estimation
- Summation of spectra, etc.
- Multiplot
(presentation software):
- Overlay of spectra
- Display of spectra in series
- Adaptation of the gain
- X-Axis title selection: data
points, field or time
- Export of data in ASCII files,
etc.
|
Computer System
 |
 |
 |
| Processor |
|
from Pentium II upwards |
| Operating system |
|
Microsoft Windows 98 SE, 2000
und XP |
| Interface |
|
RS232 or USB |
Microwave Bridge
 |
 |
 |
| Microwave power |
|
100 µW -
50mW |
| Resonator |
|
rectangular TE102 |
| Microwave frequency |
|
9.3 - 9.55 GHz |
| Tuning |
|
automatic |
|
|
Electro Magnet
 |
 |
 |
| Modulation amplitude |
|
5 µT -
0.7 mT |
| Modulation frequency |
|
100 kHz |
| Magnetic field range |
|
50 - 450 mT
(optional - 50-600 mT,
water cooling required) |
| Field sweep |
|
0 - 400 mT
(optional 0 - 550mT, water cooling required) |
| Sweep time |
|
12 s - 34 min |
| Stability |
|
2 µT |
| Homogeneity |
|
±
5 µT in the sample region |
| Cooling |
|
air (optional water) |
Case
 |
 |
 |
| Size |
|
53 x 42 x 36 cm
(length x width x height) |
| Weight |
|
about 60 kg |
| Power consumption |
|
about 500 W |
|